JUMO AQUIS touch S Assembly Instructions Manual page 103

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

Step
Action
4
The device displays the current measured value of oxygen satura-
tion and the test material temperature. The setpoint value for oxy-
gen saturation of 100 %Sat is set automatically by the device. The
user does not need to make any entries here.
5
Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK"
6
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
7
Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
18 Calibrating O-DO sensors