JUMO AQUIS touch S Assembly Instructions Manual page 82

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

16 Calibrating Ci conductivity sensors
Calibration presets for calibrating the TC curve
Parameter
Temperature com-
pensation
Starting temperature
Final temperature
82
Possible settings
Selection from
analog selection
-50 to +250 °C
-50 to +250 °C
NOTE!
Calibration of the TC curve is possible only with automatic temperature acqui-
sition.
Explanation
Temperature input for automatic acquisition of the
test/sample solution temperature during the calibra-
tion
The starting and final temperatures of the range for
which a temperature coefficient curve is to be deter-
mined.
The starting temperature must be at least 20 °C low-
er than the final temperature. The reference tem-
perature for the measurement input must lie
between the starting and final temperatures, and dif-
fer from the starting and final temperatures by at
least 2 °C.