16 Calibrating Ci conductivity sensors
84
Step
Action
3
Ensure that
the sensor has been cleaned and is immersed in the test solu-
tion,
the set reference conductivity matches the conductivity value of
the test solution.
Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
The preset reference conductivity can be changed here manually
if necessary.
tap button to
change
reference
conductivity
field
4
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
5
Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.