Calibration presets for calibrating the relative cell constant
Parameter
Reference conductivi-
ty
Calibration presets for calibrating the temperature coefficient
Parameter
Temperature com-
pensation
Reference tempera-
ture
Operation temp.
15 Calibrating CR conductivity sensors
Possible settings
0 to 9999 mS/cm
Possible settings
Selection from
analog selection
-50 to +150 °C
-50 to +150 °C
Explanation
Conductivity of the reference solution
Explanation
Temperature input for automatic acquisition of the
test/sample solution temperature during the calibra-
tion
The conductivities of a sample solution at reference
temperature and working temperature are captured
during the calibration process. This yields 2 value
pairs (temperature/conductivity). These value pairs
provide the basis for calculating the temperature co-
efficient.
The operation temp. must differ from the reference
temperature by at least 5 °C.
73