JUMO AQUIS touch S Assembly Instructions Manual page 101

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

O-DO calibration default settings
Parameter
Compensation
Temperature com-
pensation
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
Possible settings
Fixed compensation temperature
Temperature input
Interface
Selection from Analog selection
18 Calibrating O-DO sensors
Explanation
Fixed compensation temperature:
Compensation with a fixed temperature
value that is entered in the configuration
item "Compensation temperature".
Temperature input: The integrated
temperature probe of the O-DO sensor
supplies the compensation tempera-
ture.
Interface: The AQUIS touch S trans-
mits the compensation temperature to
the sensor electronics via the interface.
The source of the compensation tem-
perature is set in the configuration point
"Compensation temperature".
Temperature input for automatic acqui-
sition of the test/sample solution tem-
perature during the calibration
101

Advertisement

Table of Contents
loading

Table of Contents