Calibrating The Temperature Coefficient (Tc) - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

16.3.2

Calibrating the temperature coefficient (TC)

16 Calibrating Ci conductivity sensors
Step
Action
1
Start calibration of the temperature coefficient.
Device menu > Calibration > Select Ci analysis input or universal
input > TC calibration
2
Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
for Ci optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" operat-
ing mode, continue with step 4
3
Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
tap button to
enter measuring
field
85

Advertisement

Table of Contents
loading

Table of Contents