JUMO AQUIS touch S Assembly Instructions Manual page 36

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

10 Configuration
For digital JUMO tecLine sensors only
Configuration point
Measuring range
Free chlorine pH-dependent
Open chlorine measurement
Total chlorine
Chlorine dioxide
Chlorine dioxide tenside-neutral
Bromide
Free chlorine pH-neutral
Ozone tenside-
dependent
Ozone tenside-
neutral
Peracetic acid
Hydrogen peroxide
Decimal place
Filter Time Constant
Start of display range
Display range end
pH-compensated chlorine measure-
ment
36
Selection/
setting option
2 ppm
20 ppm
2 ppm
20 ppm
200 ppm
10 ppm
20 ppm
2 ppm
10 ppm
200 ppm
2000 ppm
20000 ppm
20000 ppm
20 %
Auto
fixed comma format
0 to 25 s
a
0 to 20000
a
0 to 20000
On
Off
Explanation
Setting of the measuring range for
the various measurands of digital
JUMO tecLine sensors
This is where you set the measuring
range of your digital JUMO tecLine
sensor according to the order details
for the sensor.
Decimal points in the display
Optimization of measured value up-
dating in the JUMO digiLine electron-
ics
The higher the value of the filter time
const., the slower the measured val-
ue is updated.
Upper/lower limit for labeling the
scale when displaying measured val-
ues, e.g. in recorder diagrams and
bar graphs
available only for pH-dependent
measurements of free chlorine:
Activation/deactivation of the pH-
compensation for the free chlorine
measurement