Measurement/Signal Input - JUMO LOGOLINE 500d Operating Manual

Pen recorder with text printing and led dot matrix display
Table of Contents

Advertisement

8.4.2 Measurement/signal input

The input signal/sensor type is selected
and the measurement ranges as well as the
filter time constants for each input channel
are defined.
If customized tables are used for
the linearisation of the input signals,
the measurement inputs and the
customized tables have to be set us-
ing the setup program.
With inputs for resistance thermom-
eter, thermocouple, current or volt-
age, the temperature values set at
the signal input are automatically
copied to the scaling after editing.
This does not apply, if linear, cus-
tomized
table 2 has been configured as line-
arisation.
Display
INPUT
1. The cold junction temperature of the internal cold junction is measured on channel 1. It is also valid for channels 2 and 3.
If one or more thermocouple inputs are configured with an internal cold junction, a thermocouple with internal cold junction
measurement or external constant cold junction temperature has to be configured for channel 1.
8 Programming
table 1
or
customized
Edit
Selection/input
E
1 INPUT
Input number of input
channel
1 TYPE: RTD
2
THERMOCOUPLE
POTENTIOMETER
RES.TRANSMITTER
VOLTAGE
CURRENT
with keys Enter
h
R
,
h
R
,
1
64
Continue with
E
2
E
3
4
5
6
7
8