14.1
Important information
14.2
General information
14.2.1
Calibration methods for ammonia sensors
Zero-point calibration
14.2.2
Calibration default settings for ammonia sensors
WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
⇨
Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
The calibration of ammonia sensors is based on measurements in ammonia-
free test solutions.
This calibration method is used to determine the ammonia zero point.
An ammonia-free test solution (e.g. water) is needed as a reference.
In the ammonia sensor calibration default settings, the zero-point calibration is
enabled and preconfigured as the only available calibration routine.
Open the calibration default settings:
Device Menu > Calibration > Select Analysis Input for pH/redox/NH
tion Presets
Sample screen:
Ammonia calibration
default settings
14 Calibrating ammonia sensors
> Calibra-
3
69