Calibrating Ammonia Sensors; Important Information; General Information; Calibration Methods For Ammonia Sensors - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

14.1

Important information

14.2

General information

14.2.1

Calibration methods for ammonia sensors

Zero-point calibration
14.2.2

Calibration default settings for ammonia sensors

WARNING!
During the calibration, the relays and analog output signals assume the states
configured for the calibration! The response of the output signals is set for each
output in its "Response at calibration" configuration point.
Chapter 10.7 "Analog outputs of base unit and optional boards", page 28
The calibration of ammonia sensors is based on measurements in ammonia-
free test solutions.
This calibration method is used to determine the ammonia zero point.
An ammonia-free test solution (e.g. water) is needed as a reference.
In the ammonia sensor calibration default settings, the zero-point calibration is
enabled and preconfigured as the only available calibration routine.
Open the calibration default settings:
Device Menu > Calibration > Select Analysis Input for pH/redox/NH
tion Presets
Sample screen:
Ammonia calibration
default settings

14 Calibrating ammonia sensors

> Calibra-
3
69