Analog Inputs; Base Unit Temperature Inputs - JUMO AQUIS touch S Assembly Instructions Manual

Modular multichannel measuring device for liquid analysis with integrated controller and paperless recorder
Hide thumbs Also See for AQUIS touch S:
Table of Contents

Advertisement

10.6

Analog inputs

10.6.1

Base unit temperature inputs

Configuration point Selection/
Description
Signal type
Connection type
Customer-specific
linearization
Start of
display range
Display range end
Decimal place
Offset
Filter Time Constant
Base unit temperature inputs: IN 4/5
Open: Device Menu > Configuration > Analog Inputs >
Temperature Inputs 1 to 2
setting option
up to 15 text characters
IN 4/5:
Pt100
Pt1000
400 Ω
4000 Ω
IN 5 :
100 kΩ
NTC 8k55
NTC 22k
Resistance potentiome-
a
ter
2-wire,
3-wire
Selection of a
linearization table
-99999 to +99999
-99999 to +99999
Auto,
Fixed comma format
b
-999 to +999
0.0 to 25.0 s
Explanation
Designation for the input
Type of sensor connected
For Pt100, Pt1000 and NTC,
corresponding linearizations are provided.
For 400 Ω, 4000 Ω and 100 kΩ, a customer-spe-
cific linearization must be configured.
For resistance transmitters
cific linearization can be configures if necessary.
"Customer-specific linearization", in this table
only for signal types Pt100, Pt1000, 400 Ω,
4000 Ω, 100 kΩ and NTC:
Connection variants for the connected resistance
thermometer
only for signal types 400 Ω , 4000 Ω , 100 kΩ
or resistance transmitters
Linearization tables contain up to 40 value pairs
in any measurement characteristic line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To cre-
ate a linearization table, you need the JUMO PC
setup program.
Operating manual B 202581.0
b
Upper/lower limit for labeling the scale when dis-
playing measured values, e.g. in recorder dia-
grams and bar graphs
b
Decimal points in the display
Correction value added to measuring value
This value can be used, for instance, to compen-
sate for measurement errors resulting from the
resistance of wires.
Optimization of measuring value updating
The higher the value of the filter time const., the
slower the measured value is updated.
10 Configuration
a
a customer-spe-
a
:
15