10 Configuration
10.11
Input alarm functions
10.11.1
Alarms for analog signals and digital sensors
Configuration point Selection/
Digital signal for hold Selection from the
Response at hold
Response at
calibration
Response at error
42
Open Temperature Inputs Alarm Configuration:
Device Menu > Configuration > Analog Inputs >
Temperature Inputs 1 to 2
Open Universal Inputs Alarm Configuration:
Device Menu > Configuration > Analog Inputs >
Universal Input 1 to 3 > Configuration
Open pH/Redox/NH
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Configuration
Open CR / Ci Analysis Inputs Alarm Configuration:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Configuration > Measuring Range 1 to 4
Open Alarm Configuration Inputs for Digital Sensors:
Device Menu > Configuration > Digital Sensors >
Digital Sensors 1 to 6 > Configuration
Open Alarm Configuration For External Analog Inputs:
Device Menu > Configuration > External Analog Inputs >
External Analog inputs 1 to 8
Open Alarm Configuration For Flow Rate:
Device Menu > Configuration > Flow Rate > Flow Rate 1 to 2
setting option
binary selector
inactive
active
hold
inactive
active
hold
normal
inactive
active
hold
Analysis Inputs Alarm Configuration:
3
Explanation
Digital signal for activating the hold function
When the hold function is activated, the alarm as-
sumes the state defined in the "Response at
hold" setting.
Specification of the alarm state when the hold
function is activated, at calibration of the particu-
lar input or at error (out of range)
inactive: alarm suppressed
active: alarm forced
hold: alarm state is maintained regardless of
alarm condition changes
normal: alarm in acc. with alarm condition