13 Calibrating redox sensors
66
Step
Action
4
Clean the Redox electrode and immerse it in the test solution. Wait
until the measured value displayed stabilizes and then confirm the
result of the measurement by pressing "OK".
5
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6
Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.