15 Calibrating CR conductivity sensors
15.3.2

Calibrating the temperature coefficient

76
Step
Action
1
Start calibration of the temperature coefficient.
Device menu > Calibration > Select CR Analysis Input > TC Cali-
bration
2
Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
For CR optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" opera-
tion mode, continue with step 4
3
Enter one of the measuring ranges 1 to 4.. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
tap button to en-
ter measuring
field