10 Configuration
Configuration point Selection/
Connection type
Customer-specific
linearization
Unit
Scale start
Scale end
Start of
display range
Display range end
Decimal place
Offset
Filter Time Constant
Ra
18
setting option
2-wire,
3-wire
Selection of a
linearization table
up to 5 text characters
b
-99999 to +99999
b
-99999 to +99999
b
-99999 to +99999
b
-99999 to +99999
Auto
Fixed comma format
b
-999 to +999
0.0 to 25.0 s
0 to 4000 Ω
Explanation
only for signal types Pt100, Pt1000, 400 Ω and
4000 Ω:
Connection variants for the connected resistance
thermometer
Linearization tables contain up to 40
value pairs in any measurement characteristic
line.
Each value pair assigns a display value
(Y-column) to a measuring value (X-column).
Up to 8 linearization tables can be stored. To cre-
ate a linearization table, you need the JUMO PC
setup program.
Operating manual B 202581.0
Process variable unit
not adjustable for pH measurement
The temperature unit is specified in the basic set-
tings.
Chapter 10.3 "Basic settings", page 10
only for standard signals:
Measured value from sensor (uncomp.) that cor-
responds to the lower limit of the standard signal
range [0 V or 0(4) mA];
Refer to the technical data for
the sensor.
only for standard signals:
only for standard signals:
Measured value from sensor (uncomp.) that cor-
responds to the upper limit of the standard signal
range [10 V or 20 mA]
Refer to the technical data for
the sensor.
Upper/lower limit for labeling the scale when dis-
playing measured values, e.g. in recorder dia-
grams and bar graphs
Decimal points in the display
only for temperature and
conductivity measurement:
Correction value added to measured value
Optimization of measured value updating
The higher the value of the filter time const., the
slower the measured value is updated.
only for IN 11/12: Resistance value between the
slider (S) and start point (A) in a resistance po-
tentiometer when the slider is at the start point