13.3
Redox calibration routines
13.3.1
Zero-point calibration
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
⇨
Chapter 8.2.1 "Log-on/Log-out", page 91
NOTE!
Digital sensors need to be linked before they can be calibrated?
⇨
Chapter 8.2.7 "Digital sensors", page 97
Step
Action
1
Ensure that
•
the calibration default settings are correct,
•
"mV" is set as the Redox unit in the configuration of the Redox
measurement input.
⇨
Chapter 13.2.2 "Calibration default settings for Redox sensors",
page 64.
⇨
Chapter 10.6.4 "pH/Redox/NH3 analysis inputs", page 21
2
Start the Zero-point calibration.
for redox sensors on analysis inputs:
Device Menu > Calibration > Select Analysis Input for pH/redox/
NH
> Zero-Point Calibration
3
for redox sensors with JUMO digiLine electronics:
Device Menu > Calibration > Digital Sensor 1 to 6 > Zero-Point
Calibration
3
Check that the "Redox test solution" value displayed matches the
Redox value of the test solution.
The "Redox test solution" value is taken from the calibration default
settings. It is still possible to change this manually here.
Tap button to
change the Re-
dox value of the
test solution
manually
13 Calibrating redox sensors
65