11.1 What to do if...
Fibre pens on 0%
No recording
Fibre pen does not write - The fibre pen has not been fitted correctly
No chart feed
Measurements are not
recorded
11 Fault finding
- The value is outside the measuring range
(Out Of Range)
- The transducer has been connected incorrectly
- The signal inputs are configured incorrectly
- Sensor break
S
- Key
has been pressed
Section 6.1
- The chart cassette has not been fitted correctly
Section 9.2
Caution: Insert the chart cassette by applying light up-
wards pressure
- The take-up roll has not properly engaged with the chart
cassette
Section 9.2.1
- The end of the chart has been reached
Section 9.2.1, Section 9.2.2
- The chart speed has been programmed as 0 mm/h
Section 8.3
Section 5.4
- The fibre pen has run out of ink
- The chart cassette has not been fitted correctly
Section 9.2
- The chart has not been fitted correctly
- The chart speed has been programmed as 0mm/h
- The pin roller does not engage with the chart perforations
- The tear-off edge in not correctly engaged
- The pen recorder is in the STOP status
- Check that the connecting terminals are tightened properly
- Check the supply
- Check the input configuration (range)
- Check the transducers and their cables, measure them
where appropriate
51