Prerequisites
Fig. 7
Installing/ removing the C-arm covers
WARNING
CAUTION
•
Remove the 4 mounting screws (Fig. 2). Hold the covers firmly while removing them.
•
Install them following the same procedure in the reverse order.
Monitor trolley
Removing the covers
•
Remove the upper back cover plate of the logbook compartment.
•
Remove the protective conductor.
•
Remove the lateral Allen screws from the lower back cover.
•
Remove the protective conductor from the cover.
Siemens AG
Medical Solutions
Risk of injuries on mechanical parts!
If not observed, slight to medium injuries, especially to the hands,
can occur.
Observe the notes concerning the C-arm carbon fiber structure.
Risk of injury!
Noncompliance can lead to light to medium injuries and/or
property damage.
The C-arm covers contain counterbalance weights and weigh
approx. 6 kg.
Hold the covers firmly when installing or removing them.
Before the C-arm cover will be loosened, the C-arm must be
moved to its middle orbital position.
If the C-arm covers are loosened, the C-arm may not be moved to
its orbital end positions (I.I. or SIREPHOS at the orbital end posi-
tion).
SPR2-230.061.01
Rev. 13
Page 15 of 18
12.04
CS PS 24
1 - 15
1
1
SIREMOBIL Iso-C
System Manual