Monitor Trolley - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

Prerequisites
CAUTION
Fig. 7:
Remove the 4 mounting screws
them.
Install them using the same procedure in the reverse order.

Monitor trolley

Removing the covers
Remove the upper back cover plate of the logbook compartment.
Remove the protective conductor.
Remove the lateral Allen screws from the lower back cover.
Remove the protective conductor from the cover.
Installing the covers
Begin with the lower back cover.
Reattach the protective conductor for the lower cover.
Install the cover and tighten the screws.
Reattach the protective conductor for the upper back cover plate.
© Siemens, 2008
For internal use only
Risk of injury!
Noncompliance can lead to minor to medium injury and/or
property damage.
The C-arm covers contain counterbalance weights and weigh
approx. 6 kg.
Hold the covers firmly when installing or removing
them.
Before the C-arm cover is loosened, the C-arm must be
moved to its middle orbital position. If the C-arm cov-
ers are loosened, the C-arm may not be moved to its
orbital end positions (I.I. or SIREPHOS at the orbital
end position).
(Fig. 7 / p.
SPR2-230.840.01.01.02
08.08
23). Hold the covers firmly while removing
Page 23 of 148
CS PS SP
23
0
SIREMOBIL Iso-C

Advertisement

Table of Contents
loading

Table of Contents