Compression and system movements
4.1

Testing the "Compression" function

Fig. 1:
Attach the spring scale as shown in
compression plate during the following tests: Tolerance: ±1 kp.
4.1.1

Switching off

System ON
PMF
Presetting
Select any value (e.g. 6 kp) on the potentiometer
the compression unit switch off at this setting?
SIE
Max. value switch off
Select the maximum value (20 kp) with the potentiometer
Does the unit switch off when it reaches the maximum value?
Correct display
PMF
OPCOMP
Position your forearm between the object table and the compression plate and check
the OPCOMP function.
MAMMOMAT 1000/3000 Nova SPB7-230.101.02.05.02
Compression and system movements
Compression
(7/Fig. 1 / p.
Typically, the cutoff value is between 6 and 10 kp at the factory setting.
02.06
(1/Fig. 1 / p. 20)
(2/Fig. 1 / p. 20)
13).
Page 20 of 34
CS PS SP
and be careful not to damage the
and compress. Does
(2/Fig. 1 / p.
20). (20 kp).