Testing

Testing the AEC-function

AEC performance test

Testing of the AEC-function is to be performed according to the test protocol, and for the
object tables in question. (Should error codes Er013 or Er450 appear; change to AEC cal-
ibration plexi which is 1 cm thinner.) Mark the exposed films with exposure conditions and
store them together with the test protocol.
NOTE
These limits apply for film with max. incremental gamma 5.0.
NOTE
Siemens AG
Medical Solutions
Tolerance only valid for film/screen combinations listed in
Table 1 under "Checking and programming with the service PC"
on Page 11 - 1.
It is important to make sensitometer exposures on all films to be
measured.
Register 3
SPB7-230.033.11
Rev. 06
Page 1 of 4
05.06
CS PS SP
13 - 1
13
13
13
M1000/3000 Nova
Installation and Start-Up