Agilent Technologies E4418B User Manual page 148

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

3
Menu Reference
Press this key to move through all the softkeys available on a particular level of
a menu. The bottom right of the power meter display indicates the number of
pages in the menu. For example, if "1 of 2" is displayed, pressing
you to "2 of 2". Pressing
124
Min
Press this softkey to enter the lower measurement limit. The
measurement units of the limits will be the same as the measurement
units in the currently selected window. On instrument preset, the default
value is either –90.000 dBm, 1 pW, –120 dB, or 100p% depending on the
windows unit. Use the
value. To confirm your choice, press the appropriate measurement units.
Min
must be less than
TTL Output
Press this softkey to access a menu which allows you to control the rear
panel TTL outputs. You can:
Switch the TTL outputs on or off.
Set the TTL output level to active high or low.
Determine whether the TTL output represents an over limit condition,
under limit condition or both.
TTL Output Off On
Press this softkey to toggle between switching the TTL output off and
on.
Limits OVER UNDER EITHER
Press this softkey to select whether the TTL output reflects an
overlimit condition, under limit condition or both.
Fail O/P HIGH LOW
Press this softkey to select whether a high or low level TTL output
represents a limits failure.
,
,
, and
Max
.
again moves you back to "1 of 2".
E4418B Power Meter User's Guide
keys to change the
moves

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents