Table 2-1 Power Sensor Connection Requirements

E4418B Power Meter User's Guide
Sensor Model
Connection Requirements
Agilent 8481A
These power sensors connect directly to the reference calibrator.
Agilent 8481H
Agilent 8482A
Agilent 8482H
Agilent N8481A
Agilent N8481H
Agilent N8482A
Agilent N8482H
Agilent E4412A
Agilent E930xA
Agilent E930xH
Agilent E9304 H18
Agilent E9304 H19
Agilent 8481D
Prior to the power meter being calibrated an Agilent 11708A 30 dB
Agilent 8484A
reference attenuator should be connected between the power sensor and
the reference calibrator. Remove this attenuator from the power sensor
input before making measurements.
This power sensor requires a 75 Ω (f) to 50 Ω (m) N-Type adapter
Agilent 8483A
(1250-0597) to connect to the reference calibrator. Remove this adapter
from the power sensor input before making measurements.
Agilent R8486A
The waveguide power sensors have two connectors. The N-Type
Agilent Q8486A
connector is the one which is used to calibrate the power meter.
Agilent V8486A
Agilent W8486A
Agilent R8486D
Agilent Q8486D
Agilent N8486AR
Agilent N8486AQ
Agilent 8481B
These power sensors are configured with an attenuator. Prior to the
Agilent N8481B
power meter being calibrated, this attenuator must be removed. The
Agilent 8482B
attenuator must be reconnected prior to making measurements.
Agilent N8482B
Agilent E930xB
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type adapter
Agilent 8485A
Agilent N8485A
(08485-60005) to connect to the reference calibrator. Remove this adapter
Agilent E4413A
before making measurements.
Agilent E9300A H24
Agilent E9300A H25
Power Meter Operation
2
33