Exponent too large
A numeric parameter was found whose exponent was larger than 32,000.
For example,
SENS:COUN 1E34000
Too many digits
A numeric parameter was found whose mantissa contained more than 255 digits, excluding
leading zeros.
Numeric data not allowed
A numeric value was received within a command which does not accept a numeric value.
For example,
MEM:CLE 24
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix.
For example,
SENS:FREQ 200KZ
Suffix too long
A suffix used contained more than 12 characters.
For example,
SENS:FREQ 2MHZZZZZZZZZZZ
Suffix not allowed
A suffix was received following a numeric parameter which does not accept a suffix.
For example,
INIT:CONT 0Hz
Character data not allowed
A discrete parameter was received but a character string or a numeric parameter was
expected. Check the list of parameters to verify that you have used a valid parameter type.
For example,
MEM:CLE CUSTOM_1
Invalid string data
An invalid string was received. Check to see if you have enclosed the character string in single
or double quotes.
For example,
MEM:CLE "CUSTOM_1
String data not allowed
A character string was received but is not allowed for the command. Check the list of
parameters to verify that you have used a valid parameter type.
For example,
LIM:STAT 'ON'
Invalid block data
A block data element was expected but was invalid for some reason. For example,
*DDT #15FET
. The 5 in the string indicates that 5 characters should follow, whereas in
this example there are only 3.
Block data not allowed
A legal block data element was encountered but not allowed by the power meter at this point.
For example
SYST:LANG #15FETC?
.
.
.
.
.
.
.
.
.
Error Messages