Press this softkey to enter the frequency of the signal to be measured on
channel A, in the range of 0.001 MHz to 999.999 GHz. The default value is
50 MHz. Use the
confirm your choice, press the appropriate frequency unit. Entering a
frequency using this softkey is only possible when you have a sensor
calibration table or frequency dependent offset table selected or you are using
an Agilent E- Series power sensor or N8480 Series power sensor (excluding
Option CFT).
Press this softkey to enter the measurement calibration factor for channel A in
the range of 1% to 150%. The default value is 100%. Use the
and
%
are using an Agilent 8480 Series power sensor or N8480 Series power sensor
with Option CFT .
The softkey menu structured to this key affects the currently selected window.
Press
meter's display. The window which is selected is highlighted by a shadowed
box and the set up you create is implemented in this window.
Press this key to access the "Meas Setup" menu. This menu allows you to set
up a variety of conditions in the currently selected measurement window, for
example: an analog or digital display, measurement limits, resolution and
channel selection.
122
Freq
,
Cal Fac
keys to change the calibration factor. To confirm your choice, press
. Entering a calibration factor using this softkey is only possible when you
to select the upper or lower measurement window on the power
Display Format
Press this softkey to access a menu which allows you to select digital or
analog displays, set the resolution and set the limits for the analog display.
Meter Dgtl Anlg
Press this softkey to toggle between an analog and digital display. Refer
to
"Selecting a Digital or Analog Display"
information.
,
, and
keys to change the frequency. To
E4418B Power Meter User's Guide
,
on page 76 for further
,
,