1 The zero set specifications are tested at 50 MHz.
E4418B Power Meter User's Guide
Power Sensor
Agilent E4412A
Agilent E4413A
Agilent E9300A
Agilent E9301A
Agilent E9304A
Agilent E9300B
Agilent E9301B
Agilent E9300H
Agilent E9301H
Agilent N8481A (exclude Option CFT)
Agilent N8481B (exclude Option CFT)
Agilent N8481H (exclude Option CFT)
Agilent N8482A (exclude Option CFT)
Agilent N8482B (exclude Option CFT)
Agilent N8482H (exclude Option CFT)
Agilent N8485A (exclude Option CFT)
Agilent N8486AR (exclude Option CFT)
Agilent N8486AQ (exclude Option CFT)
Agilent N8487A (exclude Option CFT)
Agilent N8488A (exclude Option CFT)
Agilent N8481A with Option CFT
Agilent N8481B with Option CFT
Agilent N8481H with Option CFT
Agilent N8482A with Option CFT
Agilent N8482B with Option CFT
Agilent N8482H with Option CFT
Agilent N8485A with Option CFT
Agilent N8486AR with Option CFT
Agilent N8486AQ with Option CFT
Agilent N8487A with Option CFT
Zero Set
±50 pW
±50 pW
±500 pW
±500 pW
±500 pW
±500 nW
±500 nW
±5 nW
±5 nW
1
±25 nW
1
±25 µW
1
±2.5 µW
1
±25 nW
1
±25 µW
1
±2.5 µW
1
±25 nW
1
±25 nW
1
±25 nW
1
±25 nW
1
±25 nW
1
±63 nW
1
±63 µW
1
±6.3 µW
1
±63 nW
1
±63 µW
1
±6.3 µW
1
±63 nW
1
±63 nW
1
±63 nW
1
±63 nW
Specifications
5
159