If the confidence check fails, the failure appears in the error queue. To
examine the error queue, go to the "Errors" screen (refer to

Remote Testing

To invoke the remote self test, the IEEE 488.1 compliant standard
command, *TST? is used. This command runs a full self test and returns
one of the following codes:
• 0 - no tests failed
• 1 - one or more tests failed
The remote self test consists of the following tests:
• ROM checksum
• RAM
• RAM Battery
• Display Assembly
• Calibrator
• Measurement Assembly
• Communications Assembly (Implicit)
The communications assembly is tested implicitly, in that the command
will not be accepted or return a result unless the GPIB interface is
functioning correctly.
Refer to
each individual test.
When the *TST? command is executed, the screen is cleared. As each test
takes place, the name of the test is listed on the screen. While a test is
running, the message "Testing..." appears next to the name of the test. As
each stage of the test is completed, the message "Testing..." is replaced by
either the message "Passed" or "Failed".
E4418B Power Meter User's Guide
power meter reading is made, the message "Passed" is displayed,
otherwise "Failed" is displayed.
"Test Descriptions"
on page 98, if you require a description of
Power Meter Operation
Chapter
2
4).
97