Table 5-5 Settling Time - Agilent Technologies E4418B User Manual

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

5
Specifications
For Agilent E-Series power sensors
In FAST mode, within the range –50 dBm to +17 dBm, for a 10 dB decreasing
power step, the settling time is 10 ms
1
When a decreasing power step crosses the power sensor's auto-range switch point, add 25 ms. Refer to the
relevant power sensor manual for further information.
For Agilent E- Series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:

Table 5-5 Settling Time

Number of Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
166
1
2
4
8
0.07
0.12
0.21
0.4
0.04
0.07
0.12
0.21
1
.
16
32
64
128
1
1.8
3.3
6.5
0.4
1
1.8
3.4
E4418B Power Meter User's Guide
256
512
1024
13
27
57
6.8
14.2
32

Hide quick links: