User's Guide
Data corrupt or stale
This occurs when a
FETC?
meter state has changed such that the current measurement is invalidated (for example, a
change of frequency setting or triggering conditions).
Data corrupt or stale;Please zero and calibrate Channel A
When
CAL[1|2]:RCAL
has not been zeroed and calibrated, then any command which would normally return a
measurement result (for example
message.
Data corrupt or stale;Please zero Channel A
When
CAL[1|2]:RCAL
has not been zeroed, then any command which would normally return a measurement result
(for example
,
FETC?
READ?
Data corrupt or stale;Please calibrate Channel A
When
CAL[1|2]:RCAL
has not been calibrated, then any command which would normally return a measurement
result (for example
FETC?
Data questionable;CAL ERROR
Power meter calibration failed. The most likely cause is attempting to calibrate without
applying a 1 mW power to the power sensor.
Data questionable;Input Overload
The power input to Channel A exceeds the power sensor's maximum range.
Data questionable;Lower window log error
This indicates that a difference measurement in the lower window has given a negative result
when the units of measurement were logarithmic.
Data questionable;Upper window log error
This indicates that a difference measurement in the upper window has given a negative result
when the units of measurement were logarithmic.
Data questionable;ZERO ERROR
Power meter zeroing failed. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
Hardware missing
The power meter is unable to execute the command because either no power sensor is
connected or it expects an Agilent E-Series or N8480 Series power sensor, and one is not
connected.
is attempted and either a reset has been received or the power
is set to
and the sensor currently connected to channel A
ON
FETC?
,
READ?
or
is set to
and the sensor currently connected to channel A
ON
or
) will generate this error message.
MEAS?
is set to
and the sensor currently connected to channel A
ON
,
READ?
or
MEAS?
) will generate this error message
Error Messages
MEAS?
) will generate this error
4