Agilent Technologies E4418B User Manual page 58

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

2
Power Meter Operation
34
Sensor Model
Connection Requirements
Agilent 8485D
Prior to the power meter being zeroed and calibrated, an Agilent 11708A
30 dB reference attenuator and an APC 3.5 (f) to 50 Ω (m) N-Type adapter
(08485-60005) should be connected between the power sensor and the
reference calibrator. Remove this attenuator from the power sensor input
before making measurements.
This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
Agilent 8487A
Agilent N8487A
(08487-60001) to connect to the power meter. Remove this attenuator
Agilent N8488A
before making measurements.
Agilent 8487D
Prior to the power meter being zeroed and calibrated, an Agilent 11708A
30 dB reference attenuator and an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) should be connected between the power sensor and the
reference calibrator. Remove this attenuator from the power sensor input
before making measurements.
E4418B Power Meter User's Guide