Agilent Technologies E4418B User Manual page 176

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

4
Error Messages
–310
–310
–310
–310
–310
–310
–310
–320
–330
–330
–330
152
System error;Dty Cyc may impair accuracy with ECP sensor
This indicates that the sensor connected is for use with CW signals only.
System error;Sensor EEPROM Read Failed - critical data not found or unreadable
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
System error;Sensor EEPROM Read Completed OK but optional data block(s) not found or
unreadable
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
System error;Sensor EEPROM Read Failed - unknown EEPROM table format
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp
(temperature compensation).
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
System error;Option 001 Battery charger fault
The power meter is connected to an AC power source, the battery is not fully charged and it is
not charging.
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In this instance, the power
meter detects power sensors connected to both its front and rear channel inputs.
Out of memory
The power meter required more memory than was available to run an internal operation.
Self-test Failed;
The –330, "Self-test Failed" errors indicate that you have a problem with your power meter.
Refer to
"Contacting Agilent Technologies"
faulty power meter.
Self-test Failed;Measurement Channel Fault
Refer to
"Measurement Assembly"
Assembly test.
Self-test Failed;Option 001 Battery requires replacement
The Option 001 battery is not charging to a satisfactory level and should be replaced.
on page 103 for details of what to do with your
on page 98 if you require a description of the Measurement
E4418B Power Meter User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents