Measuring Pulsed Signals; Figure 2-10 Pulsed Signal - Agilent Technologies E4418B User Manual

Power meter
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

2
Power Meter Operation

Measuring Pulsed Signals

The power meter can be used to measure the power of a pulsed signal.
The measurement result is a mathematical representation of the pulse
power rather than an actual measurement (assumes constant peak power).
The power meter measures the average power of the pulsed input signal
and then divides the measurement result by the duty cycle value to obtain
the pulse power reading. The allowable range of values is 0.001% to 100%.
The default value is 1.000%.
If duty cycle is enabled, then "Dty Cyc" is displayed.
Pulse measurements are not recommended using Agilent E4412A and E4413A power
N O T E
sensors.
An example of a pulsed signal is shown in

Figure 2-10 Pulsed Signal

68
Power
Figure 2-
B
A
E4418B Power Meter User's Guide
10.
Duty Cycle =
A
B
Time