5 To remove a sensor calibration table entry, use the
You can only edit the table name, you cannot delete it.
N O T E
6 When you have completed editing the sensor calibration table, press
Ensure that the frequency points you use cover the frequency range of the signals you
N O T E
want to measure. If you measure a signal with a frequency outside the frequency range
defined in the sensor calibration table, the power meter uses the highest or lowest
frequency point in the sensor calibration table to calculate the calibration factor.
E4418B Power Meter User's Guide
keys to select the entry, press
equivalent calibration factor is also removed and vice versa.
to return to the "Sensor Tables" screen.
Power Meter Operation
. If you delete the frequency, the