–330
–330
–330
–330
–330
–330
–330
–350
–361
–362
–363
–410
User's Guide
Self-test Failed;RAM Battery Fault
Refer to
"RAM Battery"
on page 98 if you require a description of the battery test.
Self-test Failed;Calibrator Fault
Refer to
"Calibrator"
on page 99 if you require a description of the calibrator test.
Self-test Failed;ROM Check Failed
Refer to
"ROM Checksum"
Self-test Failed;RAM Check Failed
Refer to
"RAM"
on page 98 if you require a description of the RAM test.
Self-test Failed;Display Assy. Fault
Refer to
"Display"
on page 99 if you require a description of the Display test.
Self-test Failed;Confidence Check Fault
Refer to
"Confidence Check"
Self-test Failed;Serial Interface Fault
Refer to
"Serial Interface"
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity cannot be
guaranteed.
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity cannot
be guaranteed.
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The output
buffer is cleared when power has been off or after
executed.
on page 98 if you require a description of the ROM Checksum test.
on page 96 if you require a description of this test.
on page 99 if you require a description of this test.
*RST
Error Messages
(reset) command has been
4