Zeroing and Calibrating Using TTL Inputs

You can use the TTL inputs on the rear panel Rmt I/O port to initiate
zero and calibration cycles on the power meter. The connector is an RJ- 45
Series shielded modular jack with the TTL input pins connected as shown
in

Figure 2-3 Rmt I/O Port TTL Inputs

The TTL inputs are active low and control the zero and calibration functions
as shown in

Table 2-2 TTL Inputs Control Logic

E4418B Power Meter User's Guide
Figure 2-
3.
1 2 3 4 5 6 7 8
Table 2- 2
Input 1
Input 2
1
1
1
0
0
1
0
0
Pin Number
Connection
1
none
2
Ground
3
Upper Window TTL Output
4
Lower Window TTL Output
5
TTL Input 1
6
TTL Input 2
7
Ground
8
Ground
Operation
None
CAL
ZERO
CAL
Power Meter Operation
2
35