Power Meter Operation
2

Figure 2-5 "Edit Cal Screen"

The power meter is shipped with a set of predefined sensor calibration tables.
The data in these tables is based on statistical averages for a range of Agilent
Technologies power sensors.
These power sensors are:
1
• DEFAULT
• HP 8481A
2
• HP 8482A
• HP 8483A
• HP 8481D
• HP 8485A
• HP R8486A
• HP Q8486A
• HP R8486D
• HP 8487A
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 which do not contain any data when the power meter is shipped
from the factory.
1 DEFAULT is a sensor calibration table in which the reference calibration factor and calibration factors are 100%.
This sensor calibration table can be used during the performance testing of the power meter.
2 The Agilent 8482B and Agilent 8482H power sensors use the same data as the Agilent 8482A.
E4418B Power Meter User's Guide
47