Menu Reference
130
IBFull - Sets the RTS signal line high while the receiver buffer
can accept more data and sets RTS low when the data buffer is
full. The transmitter is inhibited when CTS is low.
DTR/DSR
Press this softkey to obtain a pop- up window which allows you
to select one of the following:
OFF - Sets the DTR signal line permanently low.
ON - Sets the DTR signal line permanently high.
IBFull - Sets the DTR signal line high while the receiver buffer
can accept more data and sets DTR low when the data buffer
is full. The transmitter is inhibited when DSR is low.
This key is greyed out if RS422 interface is selected.
Echo
Press this softkey to toggle Echo "On" or "Off". When Echo is
enabled, received characters are transmitted back to the sender.
Tables
Press this softkey to access a menu which allows you to select between
sensor calibration tables and frequency dependent offset tables.
Sensor Cal Tables
Press this softkey to access a menu and screen which allows you to select
and edit sensor calibration tables. Use the
through the displayed sensor calibration tables. Sensor calibration tables
are only required for the Agilent 8480 Series power sensors and N8480
Series power sensors with Option CFT.
Edit Table
Press this softkey to access a menu and screen which allows you to
modify, add and remove frequency, calibration factors and offsets
from the currently selected table and modify the table name. Use the
,
,
frequency, calibration factors and offsets. Press
Delete
as follows:
, and
keys to move between the name,
and
keys to scroll
Change Insert
,
E4418B Power Meter User's Guide