transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Dexcom G5 is used exceeds the applicable RF compliance level above, the
Dexcom G5 should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the Dexcom G5.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Guidance and Manufacturer's Declaration –

Electromagnetic Emissions

The Dexcom G5 is intended for use in the electromagnetic environment specified in the next table. The
customer or the user of the Dexcom G5 should ensure that it is used in such an environment.
Electromagnetic Emissions Specifications
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Recommended Separation Distances Between Portable
and Mobile RF Communications Equipment and the
Receiver
The receiver is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the receiver can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the receiver as recommended in the next table, according to the maximum
output power of the communications equipment. Portable and mobile RF equipment include: baby
monitors, Bluetooth wireless headsets, wireless routers, microwave ovens, laptops with internal
Wi-Fi adapters, GSM cell phones, RFID scanners and hand-held security metal detector often used by
security screeners.
Dexcom G5 Mobile System User Guide
Technical Information
Compliance
Electromagnetic Environment Guidance
The Dexcom G5 uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The Dexcom G5 is suitable for use in all establishments
including domestic and those directly connected to the
Class B
public low-voltage power supply network that supplies
buildings used for domestic purposes.