The High Glucose Alert - Dexcom G5 MOBILE User Manual

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Table 6-C. Hyperglycemic Alert and Detection Rate Evaluation in
Reference to SMBG 30 Minutes Before and After (Pediatric, Ages 2-5
Years)
Hyperglycemic
Alert Level
(mg/dL)
120
140
180
200
220
240
300
Both sets of study data are presented and are labeled as
1
(SW10505).

The High Glucose Alert

Estimates of how well the adjustable High Glucose Alert performs are presented in Tables 6-A, 6-B
and 6-C. Table 6-A represents the hyperglycemic alert evaluation within 15 minutes of the YSI value
in the adult studies. Table 6-B represents the alert evaluation within 15 minutes of the YSI value for
a subset of the pediatric population—subjects age 6 to 17 years who had YSI measurements every
15 minutes. Table 6-C represents the alert evaluation within 30 minutes of an SMBG reading for 2- to
5-year old subjects in the pediatric studies.
Dexcom G5 Mobile System User Guide
Technical Information
True
Study
Alert
1
Rate
Original
92%
Software 505
97%
Original
90%
Software 505
98%
Original
87%
Software 505
99%
Original
85%
Software 505
98%
Original
81%
Software 505
100%
Original
80%
Software 505
99%
Original
71%
Software 505
95%
False
Hyperglycemia
Alert
Detection Rate
Rate
8%
98%
3%
99%
10%
98%
2%
100%
13%
96%
1%
93%
15%
96%
2%
93%
19%
95%
0%
97%
20%
95%
1%
98%
29%
90%
5%
96%
Original
(SW10050) or
Hyperglycemia
Missed
Detection Rate
2%
1%
2%
0%
4%
7%
4%
7%
5%
3%
5%
2%
10%
4%
Software 505
311