Dexcom G5 MOBILE User Manual page 351

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Calibration Error
Device
What you see
BG Meter
Smart
Device:
In App
Receiver
Dexcom G5 Mobile System User Guide
Troubleshooting
Problem
System will
not accept
calibration if
outside of the
40-400 mg/dL
range.
System didn't
accept recent
calibration.
(Sensor Glucose
Readings
troubleshooting
in Section 19.3
for a possible
reason).
No sensor
glucose readings
will be displayed
until error is
resolved.
(Continued on next page)
What you do
See Chapter 7.
Wait until your glucose is
between 40-400 mg/dL.
Calibrate only when your BG
meter values are between
40-400 mg/dL.
See Chapter 7.
Wait 15 minutes.
Enter 1 calibration.
If error screen still appears,
enter 1 more BG meter
value.
Wait 15 minutes.
If no sensor glucose readings
appear on the display, the
sensor needs to be replaced.
Contact Technical Support
to report error (see Section
17.1).
App:
Follow same instructions.
Tap question mark to get
more information.