Table 12-A. CGM System Difference to SMBG Within CGM Glucose Ranges
(Adult)
CGM
Glucose
Study
Range
1
(mg/dL)
Original
Overall
Software 505
Original
*40-60
Software 505
Original
*61-80
Software 505
Original
81-180
Software 505
Original
181-300
Software 505
Original
301-350
Software 505
Original
351-400
Software 505
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
(SW10505).
* For CGM ≤ 80 mg/dL, the differences in mg/dL are included instead of percent differences (%).
Dexcom G5 Mobile System User Guide
Technical Information
Number
2
of Paired
CGM-SMBG
7508
2992
731
221
968
336
3141
1362
1960
826
450
161
258
86
Mean
Median
Percent
Percent
Difference
Difference
-0.4%
-1.4%
-2.6%
-2.7%
-9.3
-8.0
-10.3
-6.0
-1.0
1.0
-4.0
-2.0
1.4%
0.0%
-2.6%
-3.1%
-0.7%
-2.8%
-1.4%
-2.0%
-0.7%
-2.6%
-0.0%
0.0%
5.0%
3.0%
3.9%
3.2%
Original
Mean
Absolute
Percent
Difference
14.0%
11.3%
11.7
13.0
10.7
10.1
14.2%
11.4%
13.0%
9.5%
10.5%
8.3%
11.9%
8.1%
Software 505
(SW10050) or
Median
Absolute
Percent
Difference
11.0%
8.6%
8.0
8.0
8.0
7.0
11.0%
8.9%
10.3%
7.4%
8.6%
6.0%
8.6%
6.7%
325