Table 7-A. Percentage of System Readings
Stratified in 2-Hour Increments After Calibration (Adult)
Time From
Calibration
0-2 hours
Software 505
2-4 hours
Software 505
4-6 hours
Software 505
6-8 hours
Software 505
8-10 hours
Software 505
10-12 hours
Software 505
12-14 hours
Software 505
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
Study
of Paired
2
CGM-YSI
Original
1929
469
Original
1516
389
Original
1547
383
Original
1520
380
Original
1555
347
Original
1068
295
Original
17
0
Within YSI Values With Data
1
Percent
Percent
Within
Within
15/15%
20/20%
YSI
YSI
78%
88%
93%
97%
69%
81%
90%
97%
69%
79%
85%
91%
68%
79%
79%
90%
71%
82%
83%
92%
65%
77%
80%
90%
65%
76%
--
--
Original
(SW10050) or
Percent
Percent
Greater
Within
Than
30/30%
40/40%
YSI
YSI
96%
2%
99%
0%
91%
4%
99%
0%
91%
5%
97%
2%
92%
3%
97%
2%
92%
4%
98%
0%
91%
4%
98%
0%
82%
12%
--
--