Dexcom G5 MOBILE User Manual page 309

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Table 5-C. Hypoglycemic Alert and Detection Rate Evaluation in
Reference to SMBG 30 Minutes Before and After (Pediatric, Ages 2-5
Years)
Hypoglycemic
Alert Level
(mg/dL)
55
60
70
80
90
100
Both sets of study data are presented and are labeled as
1
(SW10505).
Low and High Glucose Alerts
The ability of the System to detect high and low glucose levels is assessed by comparing System
results to YSI results at low and high blood glucose levels and determining if the alert may have
sounded. The System and YSI values were compared by pairing the System reading that occurred
immediately after the YSI value was collected. We suggest that you ask your doctor what alert settings
would be best for you.
The Low Glucose Alert
Estimates of how well the adjustable Low Glucose Alert performs are presented in Tables 5-A, 5-B and
5-C. Table 5-A represents the hypoglycemic alert evaluation within 15 minutes of the YSI value in the
adult studies. Table 5-B represents the alert evaluation within 15 minutes of the YSI value for a subset
of the pediatric population—subjects age 6 to 17 years who had YSI measurements every 15 minutes.
Dexcom G5 Mobile System User Guide
Technical Information
True
Study
Alert
1
Rate
Original
Software 505
25%
Original
11%
Software 505
20%
Original
29%
Software 505
20%
Original
35%
Software 505
61%
Original
51%
Software 505
78%
Original
64%
Software 505
82%
False
Alert
Rate
3%
97%
75%
89%
80%
71%
80%
65%
39%
49%
22%
36%
18%
Original
Hypoglycemia
Detection Rate
57%
100%
62%
100%
77%
100%
85%
100%
89%
100%
91%
100%
Software 505
(SW10050) or
Hypoglycemia
Missed
Detection Rate
43%
0%
38%
0%
23%
0%
15%
0%
11%
0%
9%
0%