Table 2-A. Number and Percentage of YSI Values When CGM Readings
Are "LOW" or "HIGH" (Adult)
CGM
Study
Readings
Original
"LOW"
Software
505
CGM
Study
Readings
Original
"HIGH"
Software
505
Both sets of study data are presented and are labeled as
1
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
CGM-YSI
1
Pairs
n
Cumulative
Percent
n
Cumulative
Percent
CGM-YSI
> 340
1
Pairs
n
Cumulative
Percent
n
Cumulative
Percent
YSI mg/dL
< 55
< 60
66
84
42%
54%
11
16
61%
89%
YSI mg/dL
> 320
> 280
189
220
76%
89%
40
43
89%
96%
Original
< 70
< 80
123
142
79%
92%
17
18
94%
100%
> 240
238
246
96%
99%
45
45
100%
100%
Software 505
(SW10050) or
Total
≥ 80
13
155
8%
0
18
0%
Total
≤ 240
2
248
1%
0
45
0%
295