Dexcom G5 MOBILE User Manual page 325

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Table 11-A. System Agreement to SMBG within CGM Glucose Ranges
(Adult)
CGM
Glucose
Study
Range
1
(mg/dL)
Original
Overall
Software 505
Original
40-60
Software 505
Original
61-80
Software 505
Original
81-180
Software 505
Original
181-300
Software 505
Original
301-350
Software 505
Original
351-400
Software 505
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
of Paired
2
CGM-SMBG
7508
2992
731
221
968
336
3141
1362
1960
826
450
161
258
86
Percent
Percent
Within
Within
15/15%
20/20%
SMBG
SMBG
69%
81%
77%
87%
75%
84%
73%
80%
78%
86%
77%
85%
65%
78%
74%
85%
68%
81%
80%
90%
77%
88%
83%
93%
75%
85%
90%
93%
Original
(SW10050) or
Percent
Percent
Greater
Within
Than
30/30%
40/40%
SMBG
SMBG
94%
2%
96%
1%
92%
4%
87%
7%
95%
1%
95%
1%
93%
2%
96%
1%
94%
3%
97%
1%
98%
1%
99%
0%
95%
2%
98%
1%
Software 505
323