Table 3-C. Concurrence of CGM Readings and YSI Values
Pediatric
Study)
CGM
(mg/dL)
< 40
< 40
0%
40-60
0%
61-80
0%
81-120
0%
121-160
0%
161-200
0%
201- 250
0%
251- 300
0%
301- 350
0%
351- 400
0%
> 400
0%
Dexcom G5 Mobile System User Guide
Technical Information
Row Percentage of Matched Pairs in Each CGM Glucose Range
40-
61-
81-
60
80
120
0%
0%
54%
21%
58%
16%
21%
45%
30%
1%
20%
66%
0%
1%
36%
0%
0%
4%
0%
0%
1%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
YSI (mg/dL)
121-
161-
201-
160
200
250
31%
15%
0%
5%
0%
0%
4%
0%
0%
12%
1%
0%
54%
7%
1%
40%
48%
6%
9%
44%
41%
2%
7%
50%
0%
2%
18%
0%
0%
3%
0%
0%
1%
(Original
251-
301-
351-
300
350
400
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
0%
1%
0%
0%
5%
0%
0%
36%
3%
0%
59%
21%
0%
28%
50%
16%
14%
41%
36%
Number
>
of Paired
400
CGM-YSI
0%
13
0%
19
0%
76
0%
338
0%
511
0%
596
0%
658
2%
432
0%
206
2%
86
7%
70
300