Dexcom G5 MOBILE User Manual page 334

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The receiver (MT22719) is intended for use in the electromagnetic environment specified in the next
table. The customer or the user of the receiver should ensure that it is used in such an environment.
Receiver Electromagnetic Immunity Specifications
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Surge
IEC 61000-4-5
Voltage Dips, Short
Interruptions and
Voltage Variations
on Power Supply
Input Lines
IEC 61000-4-11
IEC 60601-1-11
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Dexcom G5 Mobile System User Guide
Technical Information
IEC 60601 Test
Level
± 8 kV Contact
± 15 kV Air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% U
for 1 cycle
T
0% U
for 0.5 cycle at
T
8 phase angles
70% U
(30% dip in
T
U
) for 25 cycles
T
0% U
for 250 cycles
T
30 A/m
Transmitter
Compliance
Level
± 8 kV Contact
± 15 kV Air
± 2 kV for power
supply lines
Not applicable
± 1 kV line(s) to line(s)
Not applicable
± 1 kV line(s) to line(s)
Not applicable
0% U
for 1 cycle
T
0% U
for 0.5 cycle at
T
8 phase angles
70% U
(30% dip in
T
U
) for 25 cycles
T
0% U
for 250 cycles
T
30 A/m
Electromagnetic
Environment Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
332

Advertisement

Table of Contents
loading

Table of Contents