Dexcom G5 MOBILE User Manual page 114

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Update Calibration Prompts
Once your start-up calibration is done, your update calibration schedule begins.
The steps to enter your update calibrations are the same as your initial calibration, including
entering values in only one display device. The only difference is, with update calibrations;
you enter your BG meter value just once.
Like the reminders you received with your initial calibration, if you don't enter your BG meter
values right away, the system notifies you every 15 minutes.
Update Calibration Notifications
Device
Smart Device:
Lock Screen
Smart Device:
In App
Receiver
Tap message to confirm notification on your smart device, or press select button to confirm
it on your receiver.
Sound/Vibration Prompts
In case you can't look at your screen, both the smart device and receiver provide, with the
exception of your regular 12-hour update calibration, sound/vibration prompts notifications to
let you know it's time to calibrate or if there was a system calibration error.
Dexcom G5 Mobile System User Guide
Calibration
What you see
What it means
Enter update
calibration.
If notification
doesn't go away:
• System didn't
accept calibration
• BG values are
very different
from sensor
glucose readings
What you do
Follow steps in
Section 7.6 and 7.7.
Immediately prepare
for next calibration.