Table 8-B. Sensor Stability Relative to YSI (Accuracy Over Time
(Pediatric, Ages 6-17 Years)
Day
Study
2
of
Wear
Original
Day
1
Software 505
Original
Day
4
Software 505
Original
Day
Software
7
505
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
(SW10505).
Dexcom G5 Mobile System User Guide
Technical Information
Number
Mean
of
Absolute
Paired
Percent
CGM-
Differences
YSI
1016
21.2%
740
12.7%
810
16.0%
795
8.1%
1096
15.1%
727
10.4%
Median
Percent
Absolute
Within
Percent
15/15%
Differences
YSI
15.8%
48%
8.5%
75%
13.9%
52%
6.7%
89%
11.3%
63%
8.4%
80%
Original
(SW10050) or
) -
1
Percent
Percent
Within
Within
20/20%
30/30%
YSI
YSI
61%
78%
83%
91%
66%
87%
97%
100%
76%
89%
91%
98%
Software 505
Percent
Greater
Than
40/40%
YSI
15%
4%
3%
0%
4%
1%