Dexcom G5 MOBILE User Manual page 316

Continuous glucose monitoring system
Hide thumbs Also See for G5 MOBILE:
Table of Contents

Advertisement

Table 7-B. Percentage of System Readings
Stratified in 2-Hour Increments after Calibration (Pediatric)
Time From
Calibration
0-2 hours
Software 505
2-4 hours
Software 505
4-6 hours
Software 505
6-8 hours
Software 505
8-10 hours
Software 505
10-12 hours
Software 505
12-14 hours
Software 505
CGM readings are within 40 to 400 mg/dL, inclusive.
1
Both sets of study data are presented and are labeled as
2
(SW10505).
Calibration Stability
The System must be calibrated every 12 hours. To demonstrate performance of the System over a
12-hour calibration period, Systems were evaluated to verify that performance remains consistent
over the 12-hour calibration period. Systems were evaluated in 2-hour increments after calibration.
Performance was estimated at each 2-hour interval and stratified by glucose values by calculating
the percentage of System readings within 15 mg/dL or 15%, 20 mg/dL or 20%, 30 mg/dL or 30%, 40
mg/dL or 40% and greater than 40 mg/dL or 40% of the YSI values in Tables 7-A and 7-B.
Dexcom G5 Mobile System User Guide
Technical Information
Number
Study
of Paired
2
CGM-YSI
Original
648
545
Original
649
460
Original
630
428
Original
409
325
Original
296
305
Original
253
198
Original
37
1
Within YSI Values with Data
1
Percent
Percent
Within
Within
15/15%
20/20%
YSI
YSI
65%
75%
83%
91%
51%
67%
72%
89%
51%
61%
77%
88%
52%
68%
88%
92%
53%
69%
86%
93%
58%
74%
89%
94%
32%
38%
100%
100%
Original
(SW10050) or
Percent
Percent
Greater
Within
Than
30/30%
40/40%
YSI
YSI
87%
7%
97%
1%
86%
7%
96%
2%
80%
10%
95%
2%
85%
5%
94%
3%
84%
7%
97%
1%
89%
5%
98%
0%
65%
22%
100%
0%
Software 505
314

Advertisement

Table of Contents
loading