Console Subsystem - GE Discovery NM/CT 670 Instructions Manual

Nuclear medicine imaging system
Table of Contents

Advertisement

Regulatory Clearances
Table D-3: Console Subsystem
Work Space
Requirement
Direction of service access:
front of console
Service access width: Front of
console
Head clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
Minimum Clear
Space
914 mm (36")
There are no exposed live part hazards with the cover in place. If the console
is placed under a counter, the front edge of the console must be even with
the vertical edge of the console workspace.
Note:
762 mm (30")
This is the width of the workspace in front of the equipment. A minimum of
762 mm (30 in.) or the width of the equipment, whichever is greater, is
required.
1981.2 mm (78")
This is the height of the workspace measured from the floor at the front edge
of the equipment to the ceiling or overhead obstruction(s).
A minimum of 1981.2 mm (78 in.) or the height of the equipment, whichever
is greater, is required.
D.1.2 Regulated Minimum Working Clearance by Major Subsystem
Additional Conditions
This component is typically serviced from the front with access to
the rear.
D-5
D.1 Regulatory Clearances